For R&S®ZVA50/67/110 see below
| R&S®ZVA8 | R&S®ZVA24 | R&S®ZVA40 | |
| Number of test ports | 2 or 4 | ||
| Frequency range | 300 kHz to 8 GHz | 10 MHz to 24 GHz | 10 MHz to 40 GHz |
| Number of test points per trace | 1 to 60001 | ||
| Measurement bandwidths | 1 Hz to 1 MHz (with option: up to 30 MHz) | ||
| Max. number of internal sources | 2 | 4 | 4 |
| Operating system | Windows XP Embedded | ||
| Dynamic range at 10 Hz measurement bandwidth: | |||
| Between test ports | typ. 140 dB | typ. 135 dB | typ. 140 dB |
| With direct receiver access | typ. > 145 dB | typ. > 145 dB | typ. > 150 dB |
| Power range (typ) | –45 dBm to +15 dBm | –40 dBm to +18 dBm | –40 dBm to +18 dBm |
| With optional step attenuators | Generator step attenuators extend the lower limit of the output power range by 70 dB | ||
| Measurement time per test point | < 3.5 μs (at 1 MHz measurement bandwidth) | ||
| Data transfer time (for 201 measurements points): | |||
| via IEC/IEEE bus | < 2.9 ms | ||
| via VX11 over 100 Mbit/s LAN | < 1.3 ms | ||
| via RSIB over 100 Mbit/s LAN | < 0.7 ms | ||
| Switching time between channels | < 1 ms (with no more than 2001 points) | ||
| Switching time between instrument setups | < 10 ms (with no more than 2001 points) | ||
| R&S®ZVA50 | R&S®ZVA67 | R&S®ZVA110 | |
| Number of test ports | 2 or 4 | 2 (base unit R&S®ZVA67: 4 ports) |
|
| Frequency range | 10 MHz to 50 GHz | 10 kHz to 67 GHz | 10 MHz to 110 GHz |
| Number of test points per trace | 1 to 60001 | ||
| Measurement bandwidths | 1 Hz to 1 MHz (with option: up to 30 MHz) | ||
| Max. number of internal sources | 2 | 4 | 4 (base unit R&S®ZVA67) |
| Operating system | Windows XP Embedded | ||
| Dynamic range at 10 Hz measurement bandwidth: | |||
| Between test ports | typ. 140 dB | typ. 140 dB | typ. 130 dB |
| With direct receiver access | typ. > 150 dB | typ. > 145 dB | N/A |
| Output power at test port | typ. 18 dB | typ. 18 dB | typ. 8 dB |
| Power range (typ) | –40 dBm to +15 dBm | –40 dBm to +18 dBm | –30 dBm to +8 dBm |
| With optional step attenuators | Generator step attenuators extend the lower limit of the output power range by 50 dB |
(no step attneuators) | |
| Measurement time per test point | < 3.5 μs (at 1 MHz measurement bandwidth) | ||
| Data transfer time (for 201 measurements points): | |||
| via IEC/IEEE bus | < 2.9 ms | ||
| via VX11 over 100 Mbit/s LAN | < 1.3 ms | ||
| via RSIB over 100 Mbit/s LAN | < 0.7 ms | ||
| Switching time between channels | < 1 ms (with no more than 2001 points) | ||
| Switching time between instrument setups | < 10 ms (with no more than 2001 points) | ||